Products
Toxic Gas Sensor
for chlorine gas detection of chemical industry
Application:
Widely suitable for the detection of hydrogen chloride in industrial, underground and environmental protection fieldsProduct characteristics:
High precisionDescription
MQ-E4-Cl2 is the sensor of fixed potential electrolysis type. By redox reaction, Cl2 and O₂release charge and form the current. The value of current is proportional to the concentration of Cl2, and follow the Faraday’s Law. By testing the value of current, the concentration of Cl2 is known.
Specifications
| Media | Cl2 |
| Testing range | 0-20ppm (60ppm Max.) |
| Sensitivity | -0.6~-0.95μA/ppm |
| Resolution | 0.05ppm |
| Load resistor | 33Ω |
| Response time(T90) | ≤60s |
| Stability(per month) | <2% |
| Bias voltage | 0mV |
| Output linearity | linearity |
| Zero drift(-20℃~+40℃) | ≤0.2ppm |
| Temperature range | -20℃~50℃ |
| Humidity range | 15%-90%RH |
| Life | 2 years |
| Pressure | standard air pressure±10% |
Request Consultation