Products
O2 Sensor
Application:
Used for oxygen concentration detection in industries, mines, warehousing and environmental protection.Product characteristics:
No pollutionDescription
The MQ-E3U-O2 oxygen sensor is a constant potential type sensor. Oxygen undergoes a reduction reaction on the working electrode and a corresponding oxidation reaction occurs on the counter electrode, releasing charge to form a current. The current is proportional to the oxygen concentration and follows Faraday's law. The oxygen concentration can be determined by measuring the current.
Specifications
| Item | Parameter |
| Detection gas | Oxygen(O2) |
| Measurement Range | 0~25%VOL |
| Max Range | 30%VOL |
| Sensitivity(20.9%vol.) | (80~130)uA |
| Response time(T90) | <15S |
| Repeatability | <±2% output value |
| Output linearity | Linear |
| Load resistance(recommended) | (500/1K/2K)Ω |
| Bias | -600mV |
| Zero drift(-20℃~40℃) | <1%VOL |
| Working temperature | -20℃~50℃ |
| Working humidity | 15%~90%RH |
| Working pressure | 1atm±10% |
| Lifespan | 5 years |
Request Consultation